10.2.4 BTS 150 and BTS 500
1. Plug in the BTS150/500 power cable. Turn on the chamber by pressing the power
button in the bottom right of the panel on the chamber.
2. Launch the DIFFRAC.MEASUREMENT CENTER [4] program.
3. Switch to the CONFIGURATION plugin.
4. Open Device| Goniometer| Center and insert a template by pressing the right
mouse button and selecting Import Template.
Figure10.45: Import Template button
5. In the configuration tree, select Device| NonAmbient| BTS500 and make sure the
field under “connected component” is set to /Device/ControlRack/CPUBoard