Adjustment of a Chamber on the D8
DOC-M88-EXX268_V1_12.2016 39
Figure4.2: Final position after rotation correction
3. Repeat steps Performing a rocking curve and Physical adjustment of the
chamber rotation iteratively until the angular offset is 0°±0.05° (see figure above).
4. Repeat the step Height check for a final height check.
The scans might look different and a change in the parameters can be necessary due to
different slit sizes or small differences between temperature chambers.
Rotation alignment by introducing a software offset
Instead of physically rotating the chamber body, one can introduce a deflection angle for
the chamber into the measurement software, which corrects for the chamber misalign-
ment.