Mounting the Temperature Controller into the D8 Enclosure
48 DOC-M88-EXX268_V1_12.2016
3. To remove the supporting feet: Remove the four screws in the corners of the back
side of the controller.
4. Take off the plastic end caps.
´ Now the bottom cover can be pulled off.
5. Press the central pin into the housing and take off the foot.
6. You can remove the central pins (plastics) through the vent holes.
7. Don’t forget to mount the protecting earth wire and mount the bottom cover again.
Figure5.3: Removing the supporting feet for rack mount
ð The controller is now ready for mounting into the cabinet on supporting bars.