EasyManua.ls Logo

Bruker Anton Paar - Alignment

Bruker Anton Paar
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation and Operation of the CHC plus Humidity Chamber
72 DOC-M88-EXX268_V1_12.2016
´ This may take some minutes if the water bath has to he heated or cooled for a new
sample temperature
´ The example shows a heating chamber with Z-drive. It automatically corrects the
chamber height as function of temperature (height at 25°C is 1mm).
8. If this works successfully, you can start automatic measurements as described in your
Temperature Controller Manual and in the manuals for DIFFRAC.SUITE [4] and
WIZARD [5].
9. Check box to wait for temperature and humidity reached before starting the scan.
Every time the Corio water bath was switched off, leave it switched off for at least 20
seconds and then switch on again!
Without temperature dependent height correction, peak positions might be wrong for
temperature >25°C (alignment temperature).
Meaning of the CHC plus modes:
Constant humidity: keeps the rel. humidity while changing the sample temperature.
This is a slow process because the water bath temperature has to be adapted.
Set temperature and dry air flow
Set temperature with rate
Set temperature with rate and dry air flow
Set temperature
Set temperature and humidity
7.6 Alignment
All chambers discussed in this chapter can be aligned using the general method described
in chapter Adjustment of a Chamber on the D8 [}35].

Table of Contents

Related product manuals