Installation and Operation of the DCS 350 and DHS 1100 Chambers
80 DOC-M88-EXX268_V1_12.2016
4. Insert a new temperature (°C) and a ramp rate (°C/s) in the according fields, use mode
Set Temperature with Rate (the mode Set Temperature does NOT allow to use a
ramp rate but reaches the next temperature as fast as possible, which risks
uncontrolled heating and overshoot).
5. Press Set and wait until the temperature is reached (see figures below).
´ The example shows a heating chamber with Z-drive. It automatically corrects the
chamber height as function of temperature (height at 25°C is 1mm).
Figure8.8: Setting a new temperature and heating rate with COMMANDER
Figure8.9: Selecting the temperature mode in the COMMANDER