EasyManua.ls Logo

Bruker Anton Paar - Page 85

Bruker Anton Paar
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation and Operation of the DCS 350 and DHS 1100 Chambers
DOC-M88-EXX268_V1_12.2016 85
10.Transfer these values to the CONFIGURATION and check the box for Automatic
Correction.
´ After saving the configuration, activating it and reconnecting, these values become
active. The Z-position can no longer be set independently, but it will now be set
automatically (in COMMANDER and WIZARD) according to its calibrated positions
at the working temperature. The sample height is linearly interpolated for
intermediate temperatures.
ð Basic sample alignment in direct beam is finished.
Table8.1: Z-position as function of temperature for DHS 1100 in air
Temperature [°C] Z-Position [mm]
25 -0.657
100 -0.670
200 -0.687
300 -0.718
400 -0.730

Table of Contents

Related product manuals