Appendix / Service Details
DOC-M88-EXX268_V1_12.2016 99
10 Appendix / Service Details
10.1 Vacuum Equipment
10.1.1 Programming of the Edwards High Vacuum Stage
The programming was done at Bruker already. If necessary use the following procedure to
program Edwards TIC controller, S/W issue: D39700640R. All settings will be reset by this
procedure!
Resetting the TIC to factory defaults:
1. Press once Menu
2. Press 8 times Down Service Information
3. Press once Select
4. Press 9 times DownReset TIC Default
5. Press once SelectReset
Programming the TIC for automatic start and stop of the TMP and Backing pump:
Proceeding after resetting to factory defaults, S/W issue: D39700640R
1. Press twice Down → WRG New ID
2. Press once Select → WRG and a pressure are indicated
3. Press twice UpTurbo Off
4. Press once Select → Ramp up timer
5. Press 4 times Down → System On Setup
6. Press twice Select → Turbo No
7. Press once Down → Turbo Yes