MODEL: T-Flash
REF TR 100150
Measurable temperature range:
Forehar mode: 34 ÷ 42.2° C (93.2 ÷ 108° F)
Scan mode: – 22÷ 80° C (–7.6 ÷ 176° F)
Room operating temperature/humidity:
10 ÷ 40° C (50 ÷ 104° F) /15 ÷ 85% RH
Storage temperature/humidity range:
– 20 ÷ 50° C (–4 ÷ 122° F) /15 ÷ 85% RH
Trasportation temperature/humidity
–20 ÷ 70° C /15 ÷ 95% RH
Accuracy:
Forehead:
±0.2° C (0.4° F) in the measuring range 35 ÷ 42° C (95 ÷ 107.6°
F) ±0.3° C (0.5° F) in other measuring ranges.
Scan Mode:
±0.3° C (0.5° F) in the measuring range 22 ÷ 42.2° C
(71.6 ÷ 108° F)
±4% or ±2° C (0.4° F) in other measuring ranges
Device tested in compliance with European standards:
EN 60601-1; EN 60601-1-2; EN 12470-5; EN 60601-1-11.
- Fever alarm and Memory locations under “Ear /Forehead
(temple) Mode”.
TECHNICAL SPECIFICATIONS
GB
Error
Message
Problem Solution
Please select the target
within specifications. If a
malfunction still exists,
please contact the
distributor customer
service.
(1) In Ear mode: Tempe-
rature taken is higher than
+ 42.2° C (108° F)
(2) In Scan mode: Tem-
perature taken is higher
than + 80° C (176° F).
(1) In Ear mode: Tem-
perature taken is lower than
+ 34° C (93.2° F)
(2) In Scan mode: Tem-
perature taken is lower than
–22° C (– 7.6° F)
Device cannot be powered
on to the ready stage.
Change with a new battery.