EasyManua.ls Logo

Calex PyroUSB Series - Page 12

Calex PyroUSB Series
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
Address Length (words) Description R/W
0x11 1 Maximum Temperature R
0x12 1 Filtered Temperature R
0x13 1 Unfiltered Temperature R
0x14 1 Sensor Temperature R
0x15 1 Status (bits active high):
Bit 0: Measurement error
Bit 1: Sensor temperature low
Bit 2: Sensor temperature high
Bit 3: Object temperature low
Bit 4: Object temperature high
R
0x17 1 Reflected Temperature
Not saved to non-volatile memory
Use this register when writing frequently
R/W
0x18 1 Emissivity (1 LSB = 0.0001)
Minimum 0.1000, Maximum 1.0000
Not saved to non-volatile memory
Use this register when writing frequently
R/W
Notes:
1 All temperature are in tenths of degrees C
2 With the exception of addresses 0x17 and 0x18, all write operations are saved to non-
volatile memory
3 For further information please refer to http://www.modbus.org/specs.php
4 The default Modbus slave address of the sensor is 1; it is also possible to use slave
address 255 to communicate. 1 sensor may be connected per COM port.