9.2 Error Codes
The Cardinal 190 Weight Indicator is equipped with software that indicates
when an error in the operation takes place. The following lists the error
codes displayed by the 190 along with their meaning. Should you
encounter an error code, please refer to this list for the cause.
Error
Cause (A.) and Corrective Action (B.)
ConFig
A. Calibration required.
B. Calibrate the scale with test weights.
ErrA h
A. The analog input is off scale at the upper end of the
input range.
B. Check wiring and load cell.
ErrA L
A. The analog input is off scale at the lower end of the
input range.
B. Check wiring and load cell.
Error
A. Attempt to perform an illegal function:
1. Performing a hold operation when the weight is
invalid (e.g., overcapacity).
2. Attempting to clear Tare with Gross weight greater
than zero (lft=1, usa=0).
3. Attempting to zero scale when there is motion
present or the scale is over capacity.
4. Attempting to zero scale when Gross weight is out
of zero range (trL=yES).
5. Attempting to print when scale is over capacity.
6. Attempting to do a span adjustment with
insufficient load on the scale.
7. Attempting to tare scale when there is motion
present or the Gross weight is below zero or over
capacity.
Lo bAt
A. This message is used with the battery operation and
will turn ON to indicate the battery does not have a
sufficient charge and requires recharging.
If continued use furthers drains the battery, no change
in operation will occur until just before the battery
voltage drops to a level where operation is affected.
At this level, the indicator will automatically turn off.
B. Charge the battery pack or replace it.