Model 190 Installation, Technical and Operation
8400-M022-O1 Rev J
Page 17
4. INDICATOR SETUP
4.1 Calibration Inhibit Jumper
Your Model 190 indicator has been thoroughly tested and calibrated
before being shipped to you. If you received the indicator attached to a
scale, calibration is not necessary. If the indicator is being connected to
a scale for the first time or recalibration is necessary for other reasons,
proceed as indicated.
Calibration and Setup of the indicator is accomplished entirely by the
keypad. However, it may require changing the position of the calibration
inhibit jumper depending on the method of sealing required by your local
metrology laws.
The calibration inhibit jumper (J4) is located on the main printed
circuit board and can only be accessed by removing the front panel
of the indicator.
Figure No. 9
CAL Inhibit
Jumper J4
J4