Model 190 Installation, Technical and Operation
8400-M022-O1 Rev J
4.4 Settings
IMPORTANT! Calibration and configuration parameters are not
stored in the non-volatile memory until setup is left. If power is
lost while in setup mode, any changes made will be lost and the
indicator will revert to the previous configuration.
SETUP
USA= (Domestic or International)
With SEtUP displayed, press the TARE key. The display will change
to USA=. Press the TARE key to show the current setting. If the
setting displayed is acceptable, press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle to a new setting and then press
the TARE key to save it.
trL= (4% Zero Range) = no
trL=(4% Zero Range) = yes
CAP (Capacity) = + 4% to OC
CAP (Capacity) = + 9 grads to OC
LFt= (Legal For Trade)
Press the TARE key to show the current setting. If the setting
displayed is acceptable, press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle to a new setting and then press
the TARE key to save it.
Yes = Interval Settings (Int=)
allowed are: 1, 2, 5, 10, 20, 50
no = Interval Setting (Int=) is
selectable from 1 to 99.
When both LFt=yes and USA=yes, the followings results occur:
Scale must have between 100 and 10,000 divisions
Inhibit serial data during input
trA= (Zero Tracking Range) = 0.5 or 0 to 3
trL= (4% Zero Range) = no
CAP (Capacity) = + 4% to OC
When LFt=yes and USA=no, the followings results occur:
trL= (4% Zero Range) = yes
CAP (Capacity) = + 9 grads to OC