Model 190 Installation, Technical and Operation
8400-M022-O1 Rev J
Page 24
Unit1= (Weighing Unit 1)
Press the TARE key to show the current setting. If the setting
displayed is acceptable, press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle to a new setting and then
press the TARE key to save it. Allowable settings are:
1 = lb (pounds)
2 = kg (kilograms)
3 = oz (ounces)
4 = g (grams)
int= (Interval Setting)
Press the TARE key to show the current setting.
If LFt=YES, use the Fn/ key to toggle to a new setting and
then press the TARE key to save it. Allowable settings are:
1, 2, 5, 10, 20 or 50.
If LFt=no, use the Fn/ and UNITS/ keys to enter a new
setting and then press the TARE key to save it. Allowable
settings are: 1 through 99.
When the setting displayed is acceptable, press the TARE key
again to save it.
dPP= (Decimal Point Setting)
Press the TARE key to show the current setting. If the setting
displayed is acceptable, press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle to a new setting and then
press the TARE key to save it. Allowable settings are:
0 = XXXXXX
1= XXXXX.X
2 = XXXX.XX
3 = XXX.XXX
CAP= (Capacity)
Press the TARE key to show the current setting. If the setting
displayed is acceptable, press the TARE key again to save it.
Otherwise, use the Fn/ and UNITS/ keys to enter a new
setting and then press the TARE key to save it. Allowable
capacity settings are: 1 through 999,999.