Model 190 Installation, Technical and Operation
8400-M022-O1 Rev J
Page 32
F= (Filter Level)
The filter level is a number from 1 to 99 that corresponds to the
level of filtering with 1 being the least and 99 being the greatest.
Press the TARE key to show the current setting. To accept the
setting displayed, press the TARE key again to save it.
Otherwise, use the Fn/ and UNITS/ keys to enter a new
setting (1 to 99) and then press the TARE key to save it.
b= (Break Range)
The break range is a number from 1 to 255 that corresponds to the
number of division change to break out of the filtering. Press the
TARE key to show the current setting for the break range. To
keep the displayed setting, press the TARE key. Otherwise, use
the Fn/ and UNITS/ keys to enter a new setting (1 to 255) and
then press the TARE key to save it. NOTE: Selecting 0 disables
this feature.
Sr= (Sample Rate)
Press the TARE key to show the current setting for the sample
rate. The setting displayed is the sample rate in samples per
second. Press the TARE key to save the displayed setting or
use the Fn/ and UNITS/ keys to enter a new setting (1 to 120)
and then press the TARE key to save it.
UnS= (Motion Range)
Press the TARE key to view the current setting for the range
of motion detection. If the displayed setting is acceptable, press
the TARE key to save it. Otherwise, use the Fn/ and
UNITS/ keys to enter a new range (the number of divisions of
change permitted before indicating unstable) and then press the
TARE key to save the new setting. Allowable range values
are: 0 through 99 divisions.
SC= (Stable Count)
Press the TARE key to view the current setting for the number
of consecutive stable weight readings before indicating stable
weight. This helps filter weight readings for stability when trying
to capture stable weight. If the displayed setting is acceptable,
press the TARE key to save it. Otherwise, use the Fn/ and
UNITS/ keys to enter a new setting and press the TARE key
to save the new setting. Allowable values for the stable count
are: 1 through 255.