Model 190 Installation, Technical and Operation
8400-M022-O1 Rev J
Page 46
CAL3= Third Calibration Weight
1. The display will show CAL3=. This is the second of two
calibration weights. This weight could be ZERO (NO LOAD)
or the TEST WEIGHTS (TEST LOAD).
2. Press the TARE key to view the current setting.
3. If the second calibration weight is to be ZERO (NO LOAD),
press the TARE key.
4. If the first calibration weight is to be the TEST WEIGHTS
(TEST LOAD), use the UNITS/ and Fn/ keys to input the
value of the test weights.
5. Place the weights on scale platform, then press TARE key.
6. Starting at the left and proceeding right, a series of dashes
will appear on the display.
7. Next, starting at the left and proceeding right, the dashes will
disappear, after which the display will show: CAL4=.
CAL4= Fourth Calibration Weight
1. The display will show CAL4=. This is the fourth of the
calibration weights. This weight could be ZERO (NO LOAD) or
the TEST WEIGHTS (TEST LOAD). If the fourth calibration
weight is not going to be used, press the UNITS/ key to skip
CAL4= and advance to SetgC? prompt. Otherwise, proceed
to the next step.
2. Press the TARE key to view the current setting.
3. If the fourth calibration weight is to be ZERO (NO LOAD),
press the TARE key.
4. If the fourth calibration weight is to be the TEST WEIGHTS
(TEST LOAD), use the UNITS/ and Fn/ keys to input the
value of the test weights.
5. Place the weights on scale platform, then press TARE key.
6. Starting at the left and proceeding right, a series of dashes
will appear on the display.
7. Next, starting at the left and proceeding right, the dashes will
disappear, after which the display will show: CAL5=.