Model 190 Installation, Technical and Operation
8400-M022-O1 Rev J
CAL5= – Last Calibration Weight
1. The display will show CAL5=. This is the fifth of the calibration
weights. This weight could be ZERO (NO LOAD) or the TEST
WEIGHTS (TEST LOAD). If the fifth calibration weight is not going
to be used, press the UNITS/ key to skip CAL5= and advance to
SetgC? prompt. Otherwise, proceed to the next step.
2. Press the TARE key to view the current setting.
3. If the fifth calibration weight is to be ZERO (NO LOAD), press the
TARE key.
4. If the fifth calibration weight is to be the TEST WEIGHTS (TEST
LOAD), use the UNITS/ and Fn/ keys to input the value of the
test weights.
5. Place the weights on the scale platform, then press the TARE key.
6. Starting at the left and proceeding right, a series of dashes will
appear on the display.
7. Next, starting at the left and proceeding right, the dashes will
disappear, after which the display will show: SetgC?.