212/212X Installation & Technical
8200-M639-O1 Rev D 09/14
Page 107
DFC Weighing Procedure (Load Mode):
1. Verify the scale is in Load mode (the Unload annunciator will be
turned off). You can toggle back and forth between Load and
Unload mode by pressing the LOAD/UNLOAD key.
2. With the scale empty, verify the scale is showing 0 units Gross. If not
in Gross mode, press the 0 key then the TARE key to clear the stored
tare weight and the indicator will display Gross weight.
3. If the scale is showing a weight different than 0, press the
ZERO/CLEAR key to zero the scale.
4. If you wish to use a tare weight, place a tare weight on the scale then
press the TARE key; Or if you know the tare weight, you can manually
enter it by keying in the value followed by the TARE key.
5. The scale is now ready to be Loaded using the DFC. Press the F key
followed by the ID/START key to start the cycle.
6. The fill will start and proceed according to the SPEED and GATE settings
you selected.
7. If at any time, you wish to pause the fill cycle, press the HOLD key.
The indicator will display PAUSED. To restart/continue the filling
operation, press the HOLD key again to “unpause.” Or to abort the fill,
press the key.
8. The fill cycle will end once the preset fill weight is reached.
9. Remove the material from the scale. Repeat these steps for
subsequent fill cycles. If you wish to continue using the stored tare
weight, skip steps 1 to 4.
10. If you have a value set for PURGE above, the indicator will be counting
the number of fill cycles completed and once the purge count (PURGE
value) is reached, the purge weight (P-WGT) will be added to the trim
weight and flash PURGE FEEDER on the display. At this time, you would
need to manually purge your feeder onto the scale.