212/212X Installation & Technical
8200-M639-O1 Rev D 09/14
Page 49
trA= (Zero Tracking Range)
Press the ENTER key to show the current setting assigned to the
Automatic Zero Tracking Range. This is the setting in scale divisions that
will be automatically zeroed off. If the setting displayed is acceptable,
press the ENTER key again to save it. Otherwise, use the numeric keys
to enter a new setting and then press the ENTER key to save it.
Allowable settings are 0 (disables Zero Tracking), 0.5, or 1 through 9.
trL= (4% Zero Range)
Press the ENTER key to show the current setting. If the setting displayed
is acceptable, press the ENTER key again to save it. Otherwise, using the
numeric keys, 0/NO or 1/YES, enter the new setting and then press the
ENTER key to save it.
trL=yes
trL=no
4% of scale capacity
Full capacity (no limit)
PUO= (Power-Up Zero Feature)
Press the ENTER key to show the current setting. If the setting displayed
is acceptable, press the ENTER key again to save it. Otherwise, using the
numeric keys, 0/NO or 1/YES, enter the new setting and then press the
ENTER key to save it.
PUO=yes
PUO=no
Automatic Re-Zero on
Power-Up
No Re-Zero on Power-Up
td = (12 or 24 Time Format)
Press the ENTER key to show the current setting. If the setting displayed
is acceptable, press the ENTER key again to save it. Otherwise, use the
numeric keys to enter a new setting and then press the ENTER key to
save it. Allowable settings are 12 or 24.
td=12
td=24
12 hour clock
(3PM displays 3:00)
24 hour clock
(3PM displays 15:00)
NOTE: When using the 24 hour format, 12 is added to all times
after noon, i.e. 3 PM would be 1500.