F= (Filter Level)
The Filter Level is a number from 1 to 99 that corresponds to the
level of filtering with 1 the least and 99 being the greatest filtering.
Press the ENTER key to show the current setting. If the setting
displayed is acceptable, press the ENTER key again to save it.
Otherwise, use the numeric keys to enter a new setting and then
press the ENTER key to save it. Allowable settings are 1 to 99.
b= (Break Range)
The Break Range is a number from 1 to 255 that corresponds to
the number of division changes to break out of the filtering.
Press the ENTER key to show the current setting. If the setting
displayed is acceptable, press the ENTER key to save it.
Otherwise, use the numeric keys to enter a new setting and then
press the ENTER key to save it. Allowable settings are 0 to 255
with 0 disabling the Break Range feature.
Sr= (Sample Rate)
The Sample Rate is the measurement rate, in samples per second,
of the analog-to-digital converter.
Press the ENTER key to show the current setting. If the setting
displayed is acceptable, press the ENTER key to save it.
Otherwise, use the numeric keys to enter a new setting and then
press the ENTER key to save it. Allowable settings are 1 to 100.
UnS= (Motion Range)
The Motion Range is the number of divisions of change permitted
before indicating unstable weight.
Press the ENTER key to show the current setting. If the setting
displayed is acceptable, press the ENTER key to save it. Otherwise,
use the numeric keys to enter a new setting and then press the
ENTER key to save it. Allowable settings are 0 through 99.