212/212X Installation & Technical
8200-M639-O1 Rev D 09/14
Page 79
Calibration “C” Numbers
The calibration “C” numbers (C1, C2, C3 and C4) are displayed only during the
Test mode operation and are shown at the end of the test. Each number
is displayed for approximately 4 seconds, allowing you to record them.
Each number may be up to three (3) digits in length. By recording these
numbers you will be able to return the indicator to its present calibration
settings without using test weights simply by entering the “C” numbers.
1. With CAL1= displayed, press the "diamond T” TARE key.
2. At the C1= prompt, press the ENTER key to show the current value of
the C1 number.
3. If the C1= number displayed is acceptable, press the ENTER key again
to save it.
4. Otherwise, use the numeric keys to enter a new C1= number and then
press the ENTER key.
5. At the C2= prompt, press the ENTER key to show the current value of
the C2 number.
6. If the C2= number displayed is acceptable, press the ENTER key again
to save it.
7. Otherwise, use the numeric keys to enter a new C2= number and then
press the ENTER key.
8. At the C3= prompt, press the ENTER key to show the current value of
the C3 number.
9. If the C3= number displayed is acceptable, press the ENTER key again
to save it.
10. Otherwise, use the numeric keys to enter a new C3= number and then
press the ENTER key.
11. At the C4= prompt, press the ENTER key to show the current value of
the C4 number.
12. If the C4= number displayed is acceptable, press the ENTER key again
to save it.
13. Otherwise, use the numeric keys to enter a new C4= number and then
press the ENTER key.
14. The display will change to show: SIO?.