8200-M698-O1 Rev B 225 Installation and Technical
116
ACCUMULATORS, CONT.
To Zero A Single ID Accumulator:
1. With the indicator in normal operations mode, press the ACCUM key. The display will
change to show ACCUMULATOR=.
2. Press the ID key. The display will change to show Id=.
3. Enter the ID string and press the ENTER key.
4. The display will show Ref=XXXXXXXXXXXX. Note that XXXXXXXXXXXX is the
name associated with the ID.
5. Press the ENTER key.
6. The display will change to show the current tare weight stored.
7. Press the ENTER key.
8. The display will change to show the accumulator value of the ID entered or if the
accumulator value is greater than (>) 999,999,999, OVERFLOW will be displayed.
9. Press the ZERO key.
10. The accumulator value for the ID entered will be reset to zero.
11. Press the ENTER key to return to normal operation.
To Zero All Active ID Accumulators:
1. With the indicator in normal operations mode, press the ACCUM key. The display will
change to show ACCUMULATOR=.
2. Press the ID key. The display will change to show Id=.
3. Press the DELETE key.
4. The display will show Del All?NO, which asks if all active ID accumulators are to
be deleted.
5. To delete all active ID accumulators, press the YES key.
6. The display will change to Del All?YES.
7. Press the ENTER key.
8. The accumulator values for ALL active IDs will be reset to zero.
9. The indicator will return to normal operation when all active IDs have been zeroed.