8200-M698-O1 Rev B 225 Installation and Technical
31
MAIN PC BOARD, CONT.
P8 (CAL) – CALIBRATION INHIBIT JUMPER
When installed, this jumper inhibits (prevents) calibration of the 225 indicator.
When removed, CAL will be shown on the display (to indicate calibration is allowed) and
calibration of the 225 indicator can be performed.
RE-INSTALLING THE REAR PANEL
After all terminations have been made,
1. Remove the excess cable from the instrument enclosure and securely tighten each of
the cable gland connectors.
Do not over-tighten these connectors but make certain they are snug.
DO NOT USE TOOLS! Finger-tighten only!
2. Ensure any unused gland connectors are plugged and replace the rear panel.
3. Secure the rear panel with the 14 acorn nuts removed earlier.
4. Follow a diagonal pattern when tightening the acorn nuts.