8200-M698-O1 Rev B 225 Installation and Technical
46
CALIBRATION AND SETUP, CONT.
2. NSC=XXX (AUSTRALIA NSC REQUIREMENTS)
NOTE: In countries requiring the 225 to meet Australia NSC requirements, the USA=
parameter must be set for international use (USA=NO) to allow the 2. NSC=XXX
setup parameter selection.
With SETUP MENU #1 displayed the current setting for the NSC= parameter will be shown.
Note that XXX is the current value. If the setting displayed is acceptable, proceed to the
next setup parameter.
Otherwise, press the 2 key, the ENTER key, YES or NO (on the soft keys) and then the
ENTER key to save the new setting and return to SETUP MENU #1.
NSC=YES NSC=NO
Australia NSC Requirements
Enabled
Australia NSC Requirements
Disabled
3. LFT=XXX (LEGAL FOR TRADE)
With SETUP MENU #1 displayed the current setting for the LFT= parameter will be shown.
Note that XXX is the current value. If the setting displayed is acceptable, proceed to the
next setup parameter.
Otherwise, press the 3 key, the ENTER key, YES or NO (on the soft keys) and then the
ENTER key to save the new setting and return to SETUP MENU #1.
LFT=YES LFT=NO
Interval Settings (INTERVAL)
allowed are 1, 2, 5, 10, 20, 50
Interval Setting (INTERVAL) is
selectable from 1 to 99.
NOTE: If LFT=YES and USA=YES
NOTE: If LFT=NO
Division must be 100 to 10,000
Z TRACK = .5 or 0 to 3
Inhibit serial data during input
DATE = MM/DD/YY
Enables the Converted Units Net
Weight print feature
Z LIMIT = NO
CAP + 5% to OC
NOTE: If LFT=YES and USA=NO
STABLE = 1
DATE = DD/MM/YY
Z TRACK = 0.5
Z LIMIT = yes
CAP + 9 grads to OC
PT printed with tare
Lamp test on power up