8200-M698-O1 Rev B 225 Installation and Technical
48
CALIBRATION AND SETUP, CONT.
8. NO OF SCALES=X (NUMBER OF SCALES)
With SETUP MENU #1 displayed the current setting for the NO OF SCALES= parameter
will be shown. Note that X is the current value. If the setting displayed is acceptable,
proceed to the next setup parameter.
Otherwise, press the 8 key, the ENTER key and then using the numeric keys, enter the
number of scales and then press the ENTER key to save the new setting and return to
SETUP MENU #1. Allowable values are 1, 2 or 3.
NOTE: Support for more than 1 scale requires the 225-DS (Dual Scale) Card.
If 2 or 3 are entered for number of scales, the 225 will check for additional
scales. If additional scales are not found, it will display an error message and
default back to 1 scale.
9. TOTALIZE=XXX (SCALE TOTALIZER)
With the 225-DS (Dual Scale) Card installed and the number of scales set for 2 or 3, the 225
can sum the weights of two or more scales and make the total available to be displayed.
The totalizer can be referenced as if it were another scale. NOTE: The totalizer's scale
number will always be 4 regardless of the actual number of scales attached to the indicator.
With SETUP MENU #1 displayed the current setting for the TOTALIZE= parameter will
be shown. Note that XXX is the current value. If the setting displayed is acceptable,
proceed to the next setup parameter.
Otherwise, press the 9 key, the ENTER key, YES or NO (on the soft keys) and then the
ENTER key to save the new setting and return to SETUP MENU #1.
TOTALIZE=YES TOTALIZE=NO
Scale Totalizer Enabled Scale Totalizer Disabled
NOTE: All scales to be included in the totalizer must have identical settings for:
BASE UNITS, CNVT UNITS, INTERVAL, DDP, SAMPLES, STABLE
and STBL CNT. If these are not the same, an error message will be displayed.