8200-M698-O1 Rev B 225 Installation and Technical
51
CALIBRATION AND SETUP, CONT.
1. BAUD=X XXXXXX (COM0 PORT BAUD RATE)
With the COM0 SETUP MENU displayed the current setting for the BAUD= parameter
will be shown. Note that X is the current value and XXXXXX is the actual baud rate. If
the setting displayed is acceptable, proceed to the next setup parameter.
Otherwise, press the 1 key, the ENTER key and then using the numeric keys, enter the
value for the COM0 port baud rate and then press the ENTER key to save the new
setting and return to the COM0 SETUP MENU. Allowable values are 0 through 13.
0 = 300
4 = 4800 8 = 28800 12 = 115200
1 = 600 5 = 9600 9 = 38400 13 = 230400
2 = 1200 6 = 14400
10 = 57600
3 = 2400 7 = 19200 11 = 76000
2. PARITY=X XXXX (COM0 PARITY RATE)
With the COM0 SETUP MENU displayed the current setting for the PARITY=
parameter will be shown. Note that X is the current value and XXXX is the description.
If the setting displayed is acceptable, proceed to the next setup parameter.
Otherwise, press the 2 key, the ENTER key and then using the numeric keys, enter the
value for the COM0 parity rate and then press the ENTER key to save the new setting
and return to the COM0 SETUP MENU. Allowable values are 0, 1 or 2.
0 = NONE (No Parity) 1 = Odd Parity 2 = Even Parity
3. BITS=X (COM0 DATA BITS)
With the COM0 SETUP MENU displayed the current setting for the BITS= parameter
will be shown. Note that X is the current value. If the setting displayed is acceptable,
proceed to the next setup parameter.
Otherwise, press the 3 key, the ENTER key and then using the numeric keys, enter the
value for the COM0 data bits and then press the ENTER key to save the new setting
and return to the COM0 SETUP MENU. Allowable values are 7 or 8.
4. STOPS=X (COM0 STOP BITS)
With the COM0 SETUP MENU displayed the current setting for the STOPS=
parameter will be shown. Note that X is the current value. If the setting displayed is
acceptable, proceed to the next setup parameter.
Otherwise, press the 4 key, the ENTER key and then using the numeric keys, enter the
value for the COM0 stop bits and then press the ENTER key to save the new setting
and return to the COM0 SETUP MENU. Allowable values are 1 or 2.