8200-M698-O1 Rev B 225 Installation and Technical
68
CALIBRATION AND SETUP, CONT.
SETUP SCALE
With SETUP MENU #1 displayed, press NEXT (Navigation Keys Down Arrow) to proceed to
SETUP MENU #2.
SETUP MENU #2
1. SERIAL
2. DIO
3. PRINT TABS
4. SETUP SCALE 1
5. CALIBRATE 1
Enter Selection: 0 ` NEXT ^ PREV
4. SETUP SCALE 1
With the SETUP MENU #2 displayed, press the 4 key and then the ENTER key to proceed
to SCALE 1 SETUP MENU #1.
SCALE 1 SETUP MENU #1
1. BASE UNITS=X XX 6. SAMPLES=XXX
2. CNVT UNITS=X XX 7. STABLE=XX
3. Z TRACK=X.X 8. STBL CNT=XXX
4. Z LIMIT=XXX 9. WT INTERVALS=XX
5. PWR UP Z=XXX 10. TYPE=XXXXXXXX
Enter Selection: 0 ` NEXT ^ EXIT
1. BASE UNITS=X XX (BASE UNITS)
With SCALE 1 SETUP MENU #1 displayed the current setting for the BASE UNITS=
parameter will be shown. Note that X is the current value and XX is the description. If
the setting displayed is acceptable, proceed to the next setup parameter.
Otherwise, press the 1 key, the ENTER key and then using the numeric keys, enter the
value for the base units and then press the ENTER key to save the new setting and
return to SCALE 1 SETUP MENU #1. Allowable values are 0 through 6.
The following describes the Scale Setup for the 225. The standard 225 supports one
(1) scale. With the optional 225-DS Dual Scale card installed, it can support up to
three (3) scales. In the setup menus shown and parameters described, Scale 1 is
referenced. Substitute the scale number you are configuring for the number shown.