8200-M698-O1 Rev B 225 Installation and Technical
69
CALIBRATION AND SETUP, CONT.
2. CNVT UNITS=X XX (CONVERTED UNITS)
With SCALE 1 SETUP MENU #1 displayed the current setting for the CNVT UNITS=
parameter will be shown. Note that X is the current value and XX is the description. If
the setting displayed is acceptable, proceed to the next setup parameter.
Otherwise, press the 2 key, the ENTER key and then using the numeric keys, enter the
value for the converted units and then press the ENTER key to save the new setting
and return to SCALE 1 SETUP MENU #1. Allowable values are 0 through 6.
BASE and CONVERTED UNITS SELECTIONS
0 = none 3 = lb (pounds) 6 = t (tonnes “metric tons”)
1 = tn (tons) 4 = oz (ounces)
2 = g (grams) 5 = kg (kilograms
3. Z TRACK=X.X (ZERO TRACKING RANGE)
With SCALE 1 SETUP MENU #1 displayed the current setting for the Z TRACK=
parameter will be shown. Note that X.X is the current value. If the setting displayed is
acceptable, proceed to the next setup parameter.
Otherwise, press the 3 key, the ENTER key and then using the numeric keys, enter the
value for the zero tracking range and then press the ENTER key to save the new setting
and return to SCALE 1 SETUP MENU #1. Allowable values are 0.0 (disables Zero
Tracking), 0.5, 1.0, 2.0, 3.0, 4.0, 5.0, 6.0, 7.0, 8.0 and 9.0.
4. Z LIMIT=XXX (ZERO LIMIT RANGE)
With SCALE 1 SETUP MENU #1 displayed the current setting for the Z LIMIT=
parameter will be shown. Note that XXX is the current value. If the setting displayed is
acceptable, proceed to the next setup parameter.
Otherwise, press the 4 key, the ENTER key, YES or NO (on the soft keys) and then the
ENTER key to save the new setting and return to the
Z LIMIT=YES Z LIMIT=NO
4% (±2%) of scale capacity Full capacity (no limit)
5. PWR UP Z=XXX (POWER-UP ZERO FEATURE)
With SCALE 1 SETUP MENU #1 displayed the current setting for the PWR UP Z=
parameter will be shown. Note that XXX is the current value. If the setting displayed is
acceptable, proceed to the next setup parameter.
Otherwise, press the 4 key, the ENTER key, YES or NO (on the soft keys) and then the
ENTER key to save the new setting and return to the
PUO = YES PUO = NO
Automatic Re-Zero on Power-Up No Re-Zero on Power-Up