8200-M698-O1 Rev B 225 Installation and Technical
74
CALIBRATION AND SETUP, CONT.
4. INTERVAL=XX (INTERVAL SETTING)
With SCALE 1 SETUP MENU #2 displayed the current setting for the INTERVAL=
parameter will be shown. Note that XX is the current value for the interval (division)
setting. If the setting displayed is acceptable, proceed to the next setup parameter.
Otherwise, press the 4 key, the ENTER key and then using the numeric keys, enter the
value for the interval and then press the ENTER key to save the new setting and return
to SCALE 1 SETUP MENU #2.
If the 225 is used in a Legal For Trade application (LFT=YES), allowable values are 1,
2, 5, 10, 20 or 50.
If the 225 is NOT used in a Legal For Trade application (LFT=NO), allowable values
are 1 through 99.
5. DPP=X (DECIMAL POINT POSITION)
With SCALE 1 SETUP MENU #2 displayed the current setting for the DPP= parameter
will be shown. Note that X is the current value for the decimal point position. If the
setting displayed is acceptable, proceed to the next setup parameter.
Otherwise, press the 5 key, the ENTER key and then using the numeric keys, enter the
value for the decimal point position and then press the ENTER key to save the new
setting and return to SCALE 1 SETUP MENU #2. Allowable values are 0, 1, 2, or 3.
0 = XXXXXX 1= XXXXX .X 2 = XXXX.XX 3 = XXX.XXX
6. CAP=XXXXXXX (SCALE CAPACITY)
With SCALE 1 SETUP MENU #2 displayed the current setting for the CAP= parameter
will be shown. Note that XXXXXXX is the current value for the scale capacity. If the
setting displayed is acceptable, proceed to the next setup parameter.
Otherwise, press the 6 key, the ENTER key and then using the numeric keys, enter the
value for the scale capacity and then press the ENTER key to save the new setting and
return to SCALE 1 SETUP MENU #2. Allowable values are 1 through 9,999,999. Note
that the capacity cannot exceed 7 digits or 9,999,999.