8200-M698-O1 Rev B 225 Installation and Technical
76
CALIBRATION AND SETUP, CONT.
CALIBRATE
With the SETUP MENU #1 displayed, press NEXT (Navigation Keys Down Arrow) to
proceed to SETUP MENU #2.
SETUP MENU #2
1. SERIAL
2. DIO
3. PRINT TABS
4. SETUP SCALE 1
5. CALIBRATE 1
Enter Selection: 0 ^ PREV ` EXIT
5. CALIBRATE 1
With the SETUP MENU #2 displayed, press the 5 key and then the ENTER key to proceed
to SCALE 1 CALIBRATION MENU.
SCALE 1 CALIBRATION MENU
1. Swt= XXXXXXX 6. C1=XXX
2. Sct= XXXXXXX 7. C2=XXX
3. Zct= XXXXXXX 8. C3=XXX
4. FINE SPAN 9. C4=XXX
5. HI RES 10. CALIBRATE
Enter Selection: 0 ^ EXIT
1. Swt=XXXXXXX (SPAN WEIGHT)
2. Sct=XXXXXXX (SPAN COUNT)
3. Zct=XXXXXXX (ZERO COUNT)
The calibration parameters (Swt, Sct and Zct) represent the test load (weight) and the
analog to digital readings at zero and test load. These values are established when the
scale is calibrated. By recording the parameter values, the indicator can be returned to its
present calibration settings without using test weights simply by entering the values.
6. C1=XXX
7. C2=XXX
8. C3=XXX
9. C4=XXX
The calibration “C” numbers (C1, C2, C3 and C4) represent the present calibration
settings. Each number may be up to three digits in length. The “C” numbers are
established when the scale is calibrated. By recording these numbers, the indicator can be
returned to its present calibration settings without using test weights simply by entering the
“C” numbers.
CALIBRATION “C” NUMBERS
CALIBRATION PARAMETERS