8200-M698-O1 Rev B 225 Installation and Technical
82
CALIBRATION AND SETUP, CONT.
Single-Point for Zero Only (Only Zero) Calibration
CAL 1= (FIRST CALIBRATION WEIGHT)
With the SCALE 1 CALIBRATION MENU displayed, press the 1 and 0 keys and then the
ENTER key. The display will change to show the settings for capacity, interval, decimal
point position and CAL 1=0.0. This is the first of two calibration steps.
1. Insure the scale is empty and then press the ENTER key.
2. The display will flash CALIBRATING... for a few seconds and then change to
show CAL 2=.
CAL 2= (SECOND CALIBRATION WEIGHT)
The display will show CAL 2=XXXXXXX. This is the second of two calibration steps.
1. Press the ZERO key.
2. The display will change to show the SCALE 1 CALIBRATION MENU. NOTE: If Span
Adjustment is required, proceed to the FINE SPAN ADJUSTMENT on page 70.