8200-M698-O1 Rev B 225 Installation and Technical
83
CALIBRATION AND SETUP, CONT.
Fine Span Adjustment
4. FINE SPAN (FINE SPAN ADJUSTMENT)
If Fine Span Adjustment is
desired, place a calibrated test weight on the scale and with the
SCALE 1 CALIBRATION MENU displayed, press the 4 key and then the ENTER key. The
display will change to show SCALE 1 FINE SPAN ADJUST. Note that ADJ = XXXXXXX
is the current displayed weight.
SCALE 1 FINE SPAN ADJUST
^ UP ` DOWN < EXIT
ADJ = XXXXX
Press UP (Navigation Keys UP Arrow) to increase the span.
Press DOWN (Navigation Keys Down Arrow) to decrease the span.
Press EXIT (Navigation Keys Left Arrow) to save the new setting and return to the
SCALE 1 CALIBRATION MENU.
5. HI RES (HIGH RESOLUTION ADJUSTMENT)
If High Resolution Weight adjustment is desired, place a calibrated test weight on the scale
and with the SCALE 1 CALIBRATION MENU displayed, press the 5 key and then the
ENTER key. The display will change to show SCALE 1 HIRES ADJUSTMENT. Note that
HIRES=XXXXXX is the active weight in “high resolution” mode (in 1/10 interval).
SCALE 1 HIRES ADJUSTMENT
^ UP ` DOWN < EXIT
ZERO :zero scale PRINT : print weight
HIRES=XXXXXX
Press UP (Navigation Keys UP Arrow) to increase the span.
Press DOWN (Navigation Keys Down Arrow) to decrease the span.
Press EXIT (Navigation Keys Left Arrow) to save the new setting and return to the
SCALE 1 CALIBRATION MENU.
Press the ZERO key to zero the scale.
Press the PRINT key to print the weight (followed by the text TEST).