EasyManua.ls Logo

Carel Large ROL 160-1K2 - Pressure Adjustment and Antiscalant

Carel Large ROL 160-1K2
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
ENG
Initial start-up
WTS LARGE +0300131IT rel. 1.0 - 05.09.2022
8.2.1 Increase of the working pres- sure before the membranes
If the maximum permeate quantity (Fl01*) is not reached although the recovery value cor- responds to the
specifications in the technical data, you can increase the permeate quantity by increasing the pressure before the mem-
branes.
Notice:
Do not exceed the permitted pump pressure!
This can lead to pump damage.
Select a low working pressure!
Although the diaphragms can be operated at a pressure of 5 to 12 bar, we recommend the lowest possible working
pressure (5 to 6 bar), both during commissioning and during nor- mal system operation. This allows for a lon- ger service
life of the diaphragm(s).
To increase the pump pressure, proceed as follows:
Rotary vane pump (in models from 100 l/h to 320 l/h):
» Open or close the by-pass valve (bypass) with a slotted screwdriver
* see figure previous page
Fig. 8.a
Multi-step pump (in the models from 460 l/h to 1.000 l/h):
» Adjust the ball valve downstream of the pump manually
Fig. 8.b
» Watch the pressure change on the manometer downstream of the pump and the flow rate change on the
pure water flow meter.
» Adjust until the desired flow rate is reached.
» Let the system run for 30 minutes.
» During this time, periodically check that the rated parameters (flow rates, pressures and conductivity) stabilise
within the limit values.
» Take a permeate sample (at the extraction point (E*)) and, if neces- sary, carry out a chemical-physical and
possibly bacteriological analysis to check the suitability of the water for the intended purpose.