29
CAL 10
CAL 10-1
Function: Setting Dual Range
Range of set value: 0 ~ 1
Used key Display Descriptions
DUAL- 0
Dual range function is not used.
:Save and next
Menu navigation
~ : Set value
change
:End
DUAL- 1
Dual range function is used.
Note 1. If the resolution capability is 1/10,000 or higher, "OVER" message is displayed and return to the
CAL menu mode.
CAL 10-2
Function: Setting the applied section for the Dual Range
Range of set value: 0 ~ 99999
Used key Display Descriptions
M 1000
Dual range is applied to less than 1000kg.
M 5000
Dual range is applied to less than 5,000kg.
:Save and next
Menu navigation
~ : Set value
change
:End
M 10000
Dual range is applied to less than 10,000kg.
Note 1. If the input value is greater than the maximum value, "ERR SET" message is displayed and
returned to the CAL menu mode.