EasyManua.ls Logo

CAS CI-6000A Series - 7. Installation Procedure; Load Cell and Power Connection

CAS CI-6000A Series
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15
7. How To Install
1) Load cell connection
Connect load cell connector to load cell port which is in the backside CI-6000A
* Connecting method
PIN
1 (EXC+)
2 (SEN+)
3 (EXC-)
4 (SEN-)
5 (SIG+)
6 (SIG-)
7 (SHIELD)
Note 1. In case of 4 wires L/C connect EX+ with SEN+, and connect EX- with SEN-.
Note 2. Wire color can be different depending on the load cell’s manufacturer or its model
2) Power
Adjusted to 220V 50/60Hz at factory.
(If you want to use 110V, adjust 110V/220V jump wire which is located
in the inner part of CI-6000A)