19
CAL 5
FUNCTION : Span Calibration
KEY DISPLAY DESCRIPTION
Load the weight which was set in
CAL 3 and press ENTER key.
Span calibration.
ENTER :
Span calibration
CAL 5
Analog value of
load weight state
- - -
Span calibration is finished.
You will go to micro span
calibration(CAL 6) automatically.
Note 1. If span calibration is done without any error,
you will go to micro span calibration (CAL 6) automatically.
Note 2. If the span value is low,
error message "Err 24" is displayed. Calibrate with lower resolution.
Note 3. If the span value is high, error message "Err 25" is displayed.
CAL 6
FUNCTION : Check Micro Span Calibration
KEY DISPLAY DESCRIPTION
500.0
Setting weight is shown on the
display.
ENTER : Save and
go to next menu
▽▽▽▼▽▽▽
Check the setting weight and set
SW1 DIP1 to off.
Note 1. Confirm if the displayed weight is equal to the setting weight that you have
set in CAL 3 and remove the weight from the platter.
If "0" is displayed, set SW1 DIP1 to OFF then you will go to normal mode.
Note 2. The bias is "0" when the central lamp lights up as above display.
Each lamp means the bias of -0.3, -0.2, -0.1, 0, 0.1, 0.2, 0.3 from the left lamp