63
12. Serial Interface (COM1, COM2)
COM1,COM2 Baud Rate
F30 Device ID 00 ~ 99
COM1 Transmission Mode
F31 Baud Rate 600, 1200, 2400, 4800, 9600, 19200 bps
F32 A Use of COM1 Printer, Sub-display or Computer
F33 Output Mode Stream, Stable, Data is required, Command mode
COM2 Transmission Mode
F34 Baud Rate 600, 1200, 2400, 4800, 9600, 19200 bps
F35 A Use of COM2 Printer, Sub-display or Computer
F36 Output Mode
Stream, Stable, Data is required,
Command mode
COM1,COM2 Data Format
F37 Data Out Format 22bytes and 10bytes of CAS, 18bytes of AND
①. 22 bytes of CAS
- Data bit : 8, Stop bit : 1, Parity bit : None
- Code : ASCII
- When the data is sent to computer?
Select in SET mode
■ Transmit always : 1 is selected in F33, F36.
■ Transmit when weight is stable : 2 is selected in F33, F36.
■ Transmit when data is required : 3 is selected in F33, F36.
Indicator print output format when computer transmits 1 byte of device ID to an indicator.
- Transmission data format (22 BYTE)
. . .
DATA (8byte
)
CR LF
└──┘ └──┘ │ └──────┐ │ └──┘
US(Unstable) GS(GROSS weight) Device ID Lamp condition byte Empty Unit (kg/t)
ST(Stable) NT(NET weight)
OL(Overload)