4
10. Calibration Mode
1) How to enter calibration mode
Remove the bolt on the rear panel and put the switch to CAL ON as figure.
Adjusted to CAL ON at factory.
Turn on the power while pressing the
.
Calibrate the scale.
Put the switch to CAL OFF and perform the sealing. (See page 54)
2) Available keys
key : used for moving to the next calibration menu.
~ key : used for changing preset value.
3) Calibration menu(CAL1~CAL7)
CAL 1 : Maximum capacity set
CAL 2 : Minimum division Set
CAL 3 : Setting Weight In Span Calibration
CAL 4 : Zero calibration
CAL 5 : Span calibration
CAL 6 : Check if the calibration is done properly
CAL 7 : Input weight constant calibration after selecting national code