1
CAL 1
FUNCTION : Maximum Capacity Set
RANGE 1 ~ 99,999 kg/lb
Key Display Description
▲ : Increase of no.
◀ : Shift of digit.
Enter : Store and
move to the next menu.
ti. 03
CAL 1
10000
Maximum
Capacity
Value
Program version
CAL 1 condition
10000 kg / lb
Ref 1. The maximum capacity means the maximum weight that scale can measure.
Ref 2. Do not input the resolution, there is no need to input the resolution which is
Automatically calculated.
Ref 3. If you press Enter key, it will be moved to CAL 2.
CAL 2
FUNCTION : Minimum Division Set
RANGE 0.0005 ~ 100 kg/lb
Key Display Description
▲ : Input the next division.
Enter : Store and
move to the next menu.
CAL 2
0.01
Minimum
Division
Value
CAL 2 condition
0.01 kg / lb
Ref 1. The minimum division means the value of one division.
Ref 2. External resolution is obtained by division the min. division by the
maximum capacity. Set the resolution to be within 1/10,000.
Ref 3. If you press Enter key, it will be moved to CAL 3.