CAL 5
FUNCTION : Span Calibration Function
Key Display Description
Enter : Span
calibration and
move to the next menu.
CAL 5
LOAd
Setting weight
- - - -
GOOd
Save
CAL 5 condition
Unload the weight which was set in
CAL 3.
It is displayed the setting weight.
And then, press Enter key.
Under span calibration.
Span calibration is completed.
Check whether the displayed weight
Is same with setting weight.
Calibration is completed.
Under this condition, release the load.
Press the “ENTER” key to save
the value.
Ref 1. If Span calibration is done without any error, GOOD message is displayed
The weight of setting weight is displayed on Display screen. Check the weight.
Ref 2. If the span is low, Error message (ERR 24) is displayed.
Calibrate with lower resolution.
Ref 3. After setting the exact value, remove the setting weight and Press the
"ENTER" key to save the value.
Ref 4. In case of setting F13, it can be moved to another Platform.