22
Lower the resolution.
The resolution = allowed weight/one division
Modify the allowed weight in CAL1 or modify the division in CAL2 so
that the resolution should be below 1/10,000.
Err 22
Reason ▣
The weight for span calibration is set to be lower than 10 % of the
maximum capacity of the scale.
Trouble shooting☞
Set the weight for span calibration in CAL3 to be more than 10% of
the maximum capacity.
Err 23
Reason ▣
The weight for span calibration is set to be exceeded 100 % of the
maximum capacity of the scale.
Trouble shooting☞
Set the weight for span calibration to be within the maximum capacity
of the scale in CAL 1.
Err 24
▣ Reason
The load cell output is too small at SPAN calibration.
Trouble shooting☞
Setting of current resolution is not possible due to the error in load cell.
Proceed calibration again with less resolution.
Load cell Sense Voltage
for 5V Excitation Voltage
Recommended Resolution
2 mV 1/1,000
4 mV 1/2,000
10 mV 1/5,000