BIO X6 User Manual v. 1.0 143
10.3 Manually cleaning or sterilizing the BIO X6
Always turn off and unplug the BIO X6 before cleaning it with ethanol. Wear
gloves to avoid direct contact with the bioprinter.
The BIO X6 has a chemical-resistant coating. Clean or sterilize it with >70%
(v/v) ethanol. Spray the surface and wipe it with a soft cloth.
After using ethanol, wait at least 60 seconds before starting the BIO X6. Ethanol
is flammable, and this ensures that all remaining ethanol vapor has been
ventilated. Failure to wait can cause equipment damage and personal injury.
Task
Action
Regularity
Check
Dates
Clean door
(in/out)
Clean with a smooth cloth. Do not use
harsh solvents. The glass is sensitive to
acetone and other chemicals.
Every week
Clean top plate
Clean carefully, avoid pushing
dust/particles down to the pre-filter.
Every week
Clean inside
chamber
Remove any dirt on the rail or the lead
screw with a clean, lint-free cloth or with
your finger (with gloves on). After 6
months, it can be relubricated with clear
fluoropolymer grease.
Every week
Clean printbox
Clean with a smooth cloth, avoid using
hash solvents.
Every time it
is used
Clean printbed
(top plate)
Clean the surface with a cloth and
ethanol
>70% (v/v).
Every time it
is used
Clean printbed
(lower white part)
Clean with a smooth cloth, avoid using
harsh solvents.
Every week