Stable, well-ventilated location away from direct sunlight, with a 30cm safety zone.
Ensure voltage/frequency match, fit a 5 amp fused plug wired to convention.
Locate rotor on tapered shaft, secure with washer and locking screw hand-tightened.
Minimize imbalance by loading tubes symmetrically and evenly weighted.
Qualified personnel only to use the emergency tool to release lid after rotor stops.
Wipe external surfaces with detergent. Mop spills and check for liquid ingress into the main body.
Remove locking screw, washer. Pull rotor off shaft. Clean with detergent or autoclave at 120°C.
Use up/down arrows to set RPM or Time display.
Press Function + Lid Open. Blue dot indicates RCF or RPM mode.
Use Function + Speed arrows for rates 0-9.
Store and recall settings using Function + Pulse and Time arrows.
Load samples evenly, set parameters, close lid, press Start.
Use Pulse button for short runs; time counts up while held.
OPEn indicates lid can be opened; UUAIt requires a 5-second wait.
Lld? indicates the lid needs to be closed.
tAcho? suggests checking for obstructions fouling the rotor or lid.
bAL? indicates rotor imbalance; check tube weights and loading.
rotor? indicates speed mismatch; recheck speed or contact service.
rrEc? indicates sensor or inverter issue; contact service.
Turn on buzzer, check sensors (tachometer, rotor, lid, imbalance).
Follow steps: power off, press Function, turn on, press Time up for tests.