IN VITRO
DIAGNOSTIC USE ONLY
FOR PRESCRIPTION USE ONLY
CLIA Complexity Waived for Fingerstick Whole Blood
Before you begin
Read the complete DPP
®
HIV-Syphilis Product Insert and the DPP
®
Micro Reader User Manual before
performing the test. Laboratories with a Certificate of Waiver must follow the manufacturer's
instructions for performing the test. Any modification by the user(s) to the manufacturer’s test
procedures will result in the test no longer meeting the requirements for waived classification.
•
Cover your work station with a clean, disposable absorbent workplace cover.
•
Wear disposable gloves, gown (fluid resistant or impermeable), and eye protection (goggles or face shield).
•
Treat all patient samples as potentially infectious.
•
Let the test reach room temperature (between 18 to 25
o
C or 64 to 77
o
F) before opening pouch.
•
Remove the DPP
®
HIV-Syphilis Test Device from pouch and become familiar with it.
CONTACT INFORMATION
Tel: +1 631 924 1135 or 1-844-CHEMBIO
Email: info@chembio.com
Web: www.chembio.com
•
The results of the DPP HIV-Syphilis Test Device must be interpreted using the DPP Micro Reader (Cat. No. 70-1056-0)
DPP
®
HIV-Syphilis: Quick Reference Instructions for Fingerstick Whole Blood
Sample Loop
DPP
®
Test Device Holder
DPP
®
HIV-Syphilis Test Procedure for Fingerstick Whole Blood
Before collecting the sample, write the patient
sample ID on the DPP SampleTainer Bottle with the
BLACK CAP. Remove (unscrew) the WHITE CAP
keeping the BLACK CAP screwed onto the white
part of the cap.