REVOlVING ESTABLISHMENT OF THE STANDARD UNIT WEIGHT FOR PIECE COUNTING:
This unique funetion of the MJ series is used to easily increase the number of pieces of sample used
as your piece counting standard without taking the time to count out each standard sample piece
separately. This is important because the greater the nurrber of samplepieces used to set your standard,
the more accurate and precise your piece counting will be .
+
5
20
20
res I
pes
res I
.. ·Establishthe standard numberof sample pieces and unit
weight at 5
pes, for example, by following the
procedure on the previous page .
..·Place 15 more sample pieces on the weighing pan. The
display will show that there are 20 pieces on the pan.
.. -Leave the 20 pieces on the pan and press the F key
until 20
pes appears on the screen.
.. ·Press the TARE key to memorize the new number of
sample pieces and the new unit weight .
.. ·Follow the same procedure to increase the standard
to 100 or 1000 pieces.
NOTE: The key is to let the balance do the counting for you when establishing the standard. For
example, count out 20 pieces yourself and set the standard at 20 pieces.
Use the piece counting mode to count up to 100 pieces and set the new standard at 100 pieces.
Then use the piece counting mode again to count up to 1000 pieces and set the new standard again
at 1000 pieces.