Icon displayed Condition/Cause
See page 15.
See page 15.
Lit Measurement is under way.
Blinking Pulse detected.
“Er” is displayed when the pulse value
is below 40 or above 181.
If any of the following marks is displayed, take the specied action, then
repeat the operation.
Icon displayed Condition/Cause Corrective action
• Appears when the
blood-pressure could
not be measured
accurately.
Apply cuff properly and
measure again.
The pressure is 281mmHg
or higher.
If the pressure does not
automatically decline
during measurement,
immediately press the
“START/STOP” switch
and turn the unit off.
Take the measurement
again.
Appears when the
batteries are running low.
Replace the batteries.
Detected problem in the
device.
Contact the vendor.