The indication can be set to zero many times. Sum of loads noted in the balance
memory cannot be higher than max capacity.
Betwenn following measurements do not unpluged the balance.
The balance should be switched off by the ON/OFF key. After pressing the key
again the balance is ready to work without warm stabilization.
7.1. Pipettes calibration
In specia
l function or via cooperation with special computer software PIPETTES user
can calibrate pipettes (pipettes calibration stand). Function is described in further part of
manual.
Independently on pipettes calibration for this function steam protection ring must be
installed on (see point Unpacking balance). Container of steam curtain minimizes errors of
measurement when liquid evaporates during weighing.
Before pipettes calibration pour water to steam protection ring to 2/3 height of ring.
The set is ready to use after about an hour –this time is necessary to get moisture
stability. Remember to control distilling water level in container – surface of container