- 25 -
8.2. Calibration test
Internal calibration mass is compare
d to its value in the balance memory. This preocess is
automatic. Its result is shown on the display.
Draw. 13. The calibration test
Cal. – value of internal calibration mass
Act. – result of weighing internal calibration mass
Dif – difference between two values
Return to weighing
Changes are recorded only after return to weighing mode and confirmation
the changes. Press the ESC key many times. Following question appears
on the display. Select one of the options: ENTER – confirmation / ESC –
cancel.
(see drawing. 6. Return to weighing p. 6.1.2.Return to weighing)
8.3. Manual calibration
8.3.1. Internal calibration
1.
Enter submenu P1 – Calibration.
2. Select the function 01 Internal calibration.
3. Press the RIGHT ARROW key.
4. The balance performs the calibration automatically. During this calibration do not
load the pan.
5. After this process the balance records results of the calibration in the memory and
returns to weighing mode.
- Pressing the ESC key stops the calibration process
- If during the calibration load is on the pan display show order about error. The
calibration process is stopped.
After take load off the calibration process is finished.
If the function DRH is active user cannot stop the internal calibration process.