- 27 -
- After put the weight on the pan confirm by the Enter.
- After this procedure balance returns to menu to group P1 Calibration.
- In accordance with previous point start weighing mode.
It is recomended to select external calibration mass as its mass would be about
¾ of max balance capacity.
If the DRH function is active user cannot perform the external calibration
process.
8.4. Calibration report printout
After calibrati
on user can receive the calibration report. The report can be printed on
connected printer and sent to computer or recorded in file.
P1 08 Report printout: 1: yes – report is printed
P1 08 Report printout: 0: no – report is not printed
If the parameter has the value 1 the report is generated and sent automatically.
Draw. 15. Submenu calibration
A content of report depends on setting in submenu GLP. All options with YES attribute are
printed.
Draw. 16. Submenu GLP - setting