13
Open disc completely
and turn disc on its
backside (blank side
facing up).
14
Carefully apply a drop
of QC solution (level
as defined in
11
) to the
suction point.
15
Check sample volume:
The area marked in
blue has to be filled
completely with QC
solution.
HbA1c
HbA1c
16
Press the hinge cover down
firmly to close disc. Once
closed correctly the hinge
cover cannot be opened
again.
21
Open lid by selecting
Open on the screen,
remove first HbA1c disc,
insert second HbA1c
disc, and close lid.
22
Verify that the QC solution lot
number matches the lot num-
ber displayed on the screen.
Select
to begin the mea-
surement.
23
When measurement is
complete the results of
both levels (1+2) are
displayed.
* If alarm is enabled.
24
Finish test by selecting
on the screen
and dispose of discs
appropriately.
17
Open lid, remove HbA1c
QC Info disc, insert
HbA1c disc and close lid.
18
Verify that the QC solution
lot number matches the lot
number displayed on the
screen. Select
to begin the
measurement.
19
When measurement is
complete, select to
proceed with the QC test
of the other level.
20
Open disc pouch and
remove second HbA1c
disc. Repeat procedure
shown in steps 14-16.