iv Pump Drives Operating Manual Cole-Parmer
This product is not designed for, nor intended for use in patient
connected applications; including, but not limited to, medical and
dental use, and accordingly has not been submitted for FDA
approval.
This product is not designed for, nor intended for use in
hazardous duty areas as defined by ATEX or the NEC
(National Electrical Code); including, but not limited to use with
flammable liquids. Consult the factory for products suitable for
these types of applications.
1. Read instructions before operating the unit.
2. Observe safety precautions at all times, especially when pumping
dangerous liquids.
3. If the pump runs unusually noisy or if bunching of the tubing in the
pump can be observed, make sure the tubing is clamped down tightly
and/or replace it with a new piece of tubing.
4. The Pump Drives must be well-grounded at all times.
5. The Pump Drives are equipped with a current-limiting circuit that will
slow the motor down if any of the following conditions exist:
a. Tubing that is too hard is loaded in the pump.
b. Incorrect tubing size or wall thickness is loaded in the pump.
c. Tubing is improperly loaded into the Pump Head.
6. The unit is fused and grounded to protect the operator in the event of
short circuits that could be caused by liquid entering the case.
CAUTION: Replace the fuse only with one of the same type and
rating. The fuse rating and type are stated on the rear panel.
7. The Pump Drives should not be used in outdoor or hazardous
locations.
WARNING:
Product Use
Limitation
Safety